用于动脉血压监测的柔性超声换能器阵列
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

TH871. 7

基金项目:

国家自然科学基金(61773283)项目资助


Flexible ultrasonic transducer array for arterial blood pressure monitoring
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    作为重要的生命体征之一,血压是心血管疾病风险评价指标。 中心动脉压与靶器官损害和心血管疾病有很强的病理生 理联系,动脉血压监测对心脑血管事件有较高的预测价值,可为人体心脏机能、动脉弹性表征等临床诊断提供依据。 本文提出 的超声压电换能器阵列相比于传统的刚性超声探头,具有柔性、可以贴合人体颈部的特点。 制作并对比了低频(2. 5 MHz)和高 频(4. 5 MHz)的 1×16 线性超声换能器阵列,两个阵列均在仿组织超声体模上进行了成像,穿透深度分别为 90 和 40 mm。 换能 器阵列在相控阵聚焦发射模式下已实现对颈动脉的连续血压重复检测。 与医用中心动脉压检测仪的测量结果对比表明,在静 息状态下柔性阵列与医用检测仪所测血压波形的平均相对误差为 3. 83% ,剧烈运动后平均相对误差为 5. 47% ,动脉血压波形测 量结果准确,医用价值丰富。

    Abstract:

    As one of the important vital signs, blood pressure is a risk assessment indicator for cardiovascular disease. Central arterial pressure has a strong pathophysiological relationship with target organ damage and cardiovascular disease. Arterial blood pressure monitoring has high predictive value for cardiovascular and cerebrovascular events, which can provide a basis for clinical diagnosis, such as human cardiac function and arterial elastic characterization. The proposed ultrasonic piezoelectric transducer array has the characteristics of flexibility and being able to fit the human neck compared to traditional rigid ultrasonic probes. The 1 × 16 linear ultrasonic transducer array with low frequency (2. 5 MHz) and high frequency (4. 5 MHz) is produced and compared. Both arrays are imaged on ultrasonic phantom, with penetration depths of 90 and 40 mm, respectively. The transducer array has achieved continuous and repeated blood pressure detection of the carotid artery in the phased array-focused emission mode. Compared with the measurement results of the medical central arterial pressure detector, the average relative error of the blood pressure waveform measured by the flexible array and the medical detector in the resting state is 3. 83% , and the average relative error after intense exercise is 5. 47% . The measurement results of the arterial blood pressure waveform are accurate and the medical value is rich.

    参考文献
    相似文献
    引证文献
引用本文

刘 畅,孙 贺,任佳豪,程寰宇,谢梦莹,刘 洋.用于动脉血压监测的柔性超声换能器阵列[J].仪器仪表学报,2023,44(12):225-234

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2024-02-27
  • 出版日期:
文章二维码