    摘要:

    孔口流量计是标准孔板流量计的特殊使用形式,为环境监测领域中常用流量测量元件。 孔口流量计在校准时按上游入 口压力的不同,其校准方法分为微正压和微负压两种校准工况。 两种工况均为国家标准和规范推荐方法,但校准出的流量系数 差距达 10% 。 夹具引入的附加差压和仪表流量系数线性度误差是导致两种校准工况流量系数偏差的原因。 通过 CFD 仿真研 究,设计对被校准孔口流量计上游压力影响较小的夹具形式和结构参数。 在仿真数据的基础上计算得到适用于常用管径和雷 诺数范围的孔口流量计的可膨胀修正系数公式。 运用上述成果并进行试验验证,正压和负压校准流量系数的差距减小至 ±0. 5% 以内。 校准流量范围内孔口流量计流量系数线性度也有 0. 1% 左右的提升。

    Abstract:

    The orifice flowmeter is a kind of flow measurement element which is commonly used in the field of environmental monitoring. According to the difference of upstream inlet pressure of calibrated orifice flowmeter, calibration methods of the orifice flowmeter can be categorized into two kinds of calibration conditions, which are micro positive pressure and micro negative pressure. However, both working conditions are recommended by national standards and specifications, and the deviation of the two calibration conditions is up to 10% . The calibration deviation of the two working conditions is caused by additional differential pressure value introduced by clamping device used in positive pressure condition and linearity error introduced by the lack of available expansibility factor formula. The forms and structural parameters of clamping device which is insensitive to the upstream pressure of the calibrated orifice flowmeter are designed by the computational fluid dynamics simulation method. An available expansibility factor formula for the orifice flowmeter is calculated based on the simulation data. The achieved formula is used to carry out evaluation. The deviation of flow coefficient between the micro positive pressure and micro negative pressure calibration is less than ±0. 5% . The flow coefficient linearity of the orifice flowmeter within the calibrated flow range is also improved by about 0. 1% .

    参考文献
    相似文献
    引证文献
引用本文

宋述古,刘伟光,张 强,李 忠,张晓妮.孔口流量计正压测试夹具和可膨胀系数的研究[J].仪器仪表学报,2022,43(1):19-27

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2023-02-06
  • 出版日期: