基于铯光泵磁力仪的地震地磁矢量测量系统
DOI:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

TH762

基金项目:

中国地震局地震科技星火计划项目(XH21011)、徐州市重点研发计划(社会发展)(KC20185)项目资助


Seismic geomagnetic vector measurement system based on optically pumped cesium magnetometer
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    为解决地震地磁矢量观测中质子矢量磁力仪噪声大,长期观测中存在明显漂移,磁通门磁力仪温度效应明显,保温磁房 投入大等问题,研制了基于铯光泵磁力仪的地震地磁矢量测量系统,本文详细介绍了其测量原理及仪器结构,设计了混合分量 线圈装置,并开展了稳定性分析。 研制的样机在蒙城地震台开展了观测实验。 结果表明,实验期内记录良好,采样率达到 1 Hz, 1 月份总强度、水平分量和磁偏角月噪声分别为 0. 01 nT、0. 02 nT 和 0. 003′,优于地震地磁观测基本网质子矢量磁力仪最好噪 声水平,也优于同场地两台磁通门磁力仪的噪声水平和基准网的平均噪声水平。 其实验期内水平分量和磁偏角最大漂移分别 为 1. 3 nT 和 0. 16′,小于座钟式质子矢量磁力仪基线漂移,具有良好的稳定性。

    Abstract:

    To solve the problems of high noise and obvious drift in long-term observation of proton vector magnetometer, obvious temperature effect of fluxgate magnetic field meter and large investment in thermal insulation magnetic room, a seismic geomagnetic vector measurement system based on the optically pumped cesium magnetometer is developed. In this article, the measuring principle and instrument structure are introduced in detail, the mixed-component coil device is designed, and the stability is analyzed. The prototype is tested at the Mengcheng seismic station. The results show that, during the experimental period, the instrument records well, the sampling rate of the experimental instrument reaches 1 Hz, the monthly noise of the total intensity, the horizontal component and the magnetic declination in January are 0. 01 nT, 0. 02 nT and 0. 003′, respectively, which are obviously superior to the best noise level of proton vector magnetometers of seismic geomagnetic observation basic network. They are also better than the noise level of two fluxgate magnetic field meters at the same venue and the average noise level of the seismic geomagnetic observation reference network. The maximum drift of the horizontal component and the magnetic declination are 1. 3 nT and 0. 16′, respectively, which is less than the baseline drift of the pedestal-clock proton vector magnetometer, so good stability can be obtained.

    参考文献
    相似文献
    引证文献
引用本文

居海华,张 乐,夏 忠,张 岑,应允翔.基于铯光泵磁力仪的地震地磁矢量测量系统[J].仪器仪表学报,2023,44(9):111-120

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2024-01-24
  • 出版日期: